Comparison of a new forehead reflectance pulse oximeter sensor with a conventional digit sensor in pediatric patients
John W Berkenbosch1, Joseph D Tobias
1Kosair Children's Hospital, University of Louisville, 571 S Floyd, Suite 332, Louisville, KY 40202, USA. john.berkenbosch@louisville.edu
Background:
During conditions of poor perfusion, the accuracy of conventional extremity-based pulse oximeters may be limited. Limited evidence suggests that forehead perfusion may be better preserved during such periods, but pediatric experience with newer forehead reflectance sensors is limited. We prospectively compared the accuracy of a forehead reflectance sensor, the Max-Fast, with a new-generation digit sensor in pediatric patients.
Methods:
Pediatric patients > 10 kg and who had arterial catheters were eligible for enrollment. Blood oxygen saturation was simultaneously measured with forehead and digit sensors, and compared to corresponding CO-oximetry-measured arterial oxygen saturation values (S(aO2)) taken at the same times. We used Bland-Altman analysis to calculate the bias and precision of the forehead sensor and the digit sensor relative to the S(aO2) values.
Results:
We obtained 116 sample sets from 28 patients. The S(aO2) values ranged from 84.1% to 99.2%. The bias and precision of the forehead-to-S(aO2) difference were 0.6% and 2.7%, respectively, versus 1.4% and 2.6%, respectively, for the digit-to-S(aO2) difference (p < 0.05). Bias and precision were 0.7% and 2.6% versus 1.7% and 2.3% for the forehead and digit sensors, respectively, (p < 0.05) in patients who received vasoactive medications, compared with 0.5% and 2.8% versus 1.1% and 2.8% (p = not significant), respectively, in patients who did not receive vasoactive medications.
Conclusions:
The Max-Fast sensor estimated S(aO2) as accurately as did a new-generation digit sensor in well-perfused pediatric patients.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Special considerations while measuring pulse
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...