Microfluidics in Assessing Platelet Function
Published on: November 8, 2024
Evaluation of angiographic contrast media and platelet function with impedance aggregometry and the PFA-100 'platelet
M C Dalby1, A Dewar, S J Davidson
1Department of Cardiology, Royal Brompton Hospital, London, UK. m.dalby@rbh.nthames.nhs.uk
Background:
Angiographic contrast media are used in balloon angioplasty and may influence thrombotic complications of the procedure. We studied the effect of different media on platelet aggregation in whole blood using impedance aggregometry and the PFA-100 'platelet function analyser' (Dade, USA).
Methods:
Venous blood samples from 18 healthy volunteers were split into four aliquots and mixed with 10% normal saline control, non-ionic medium (iohexol), low-molecular weight ionic medium (ioxaglate) and high-molecular weight ionic medium (diatrizoate). Samples were studied with impedance aggregometry and the PFA-100.
Results:
All media caused significant inhibition of aggregation compared with control with both methods (P<0.001). Antiplatelet potency was greatest with diatrizoate, intermediate with ioxaglate and least with iohexol with both methods (P<0.01). Electron microscopy of the PFA-100 membrane demonstrated occlusion of the experimental aperture with platelet thrombus in the control. Inhibition of platelet thrombus was seen with all media, greatest with diatrizoate, intermediate with ioxaglate and least with iohexol.
Conclusions:
The media studied significantly inhibited platelet aggregation in vitro and potency was greater with ionic than non-ionic media. These methods use a combination of shear and chemical agonist with whole blood and may reproduce in vivo arterial conditions better than other techniques.