Published on: January 2, 2012
Analysis of the gas phase of cigarette smoke by gas chromatography coupled with UV-diode array detection
Dimitris G Hatzinikolaou1, Verner Lagesson, Anastasia J Stavridou
1Institute of Biomedical Research and Biotechnology, 55 Solomou Street, 104 32 Athens, Greece.
Abstract:
A gas chromatography method, coupled with diode array photometric spectral detection in the ultraviolet region (167-330 nm), was developed for the analysis of the gas phase of cigarette smoke. The method enabled us to identify more than 20 volatiles present in the vapor phase of cigarette smoke. In that way, all major volatile organic compounds (including aldehydes, conjugated dienes, ketones, sulfides, furans, and single-ring aromatics), as well as nitric oxide (NO) and hydrogen sulfide (H(2)S), can be analyzed in a straightforward manner through a single chromatographic run of <50-min duration. The method can easily be applied by the introduction of a small volume of the gas-phase stream into the GC injection loop directly through the smoking apparatus exhaust circuit, thus providing an excellent alternative to available methods, which usually require extraction or concentration steps prior to any chromatographic analysis. Furthermore, all problems concerning aging of the gas phase are eliminated. Twelve compounds (including NO) were chosen for quantification through the use of appropriate calibration standards. Comparison of the vapor phase yields of these compounds for the reference cigarette Kentucky 1R4F with already reported data indicates that this method is very reliable as far as accuracy and reproducibility of the results are concerned. Finally, the proposed methodology was used to compare the concentration of these cigarette smoke gas-phase constituents among individual puffs.
More Related Videos
10:44Comparing the Effects of Electronic Cigarette Vapor and Cigarette Smoke in a Novel In Vivo Exposure System
Published on: May 24, 2017
09:30A Microcontroller Operated Device for the Generation of Liquid Extracts from Conventional Cigarette Smoke and Electronic Cigarette Aerosol
Published on: January 18, 2018
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...