Related Experiment Video
Updated: Aug 7, 2026

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
A validated liquid chromatographic-tandem mass spectroscopy method for the quantification of abiraterone acetate and
Vanessa Martins1, Yasmin Asad, Nicola Wilsher
1Cancer Research UK Centre for Cancer Therapeutics, The Institute of Cancer Research, 15 Cotswold Road, Sutton, Surrey SM2 5NG, UK.
Abstract:
A sensitive and selective LC-MS/MS method has been developed and validated for the quantification of abiraterone acetate and its metabolite, abiraterone (an androgen biosynthesis inhibitor) in human plasma. Analytes were extracted by SPE with cation mixed-mode polymer cartridges. Chromatography was performed on a Luna C5 5 microm, 50 mm x 2.1 mm i.d. column, using a mobile phase of 2% propan-2-ol in acetonitrile and 10mM ammonium acetate. The assay was linear from 5 to 500 nM (r(2)=0.998). The intra- and inter-day coefficients of variation were <13.9% for both analytes. This method will be applied to a clinical trial investigating the pharmacokinetics of abiraterone acetate and abiraterone in patients with prostate cancer.