Clinical use of indices determined non-invasively from the radial and carotid pressure waveforms
Audrey Adji1, Kozo Hirata, Michael F O'Rourke
1St Vincent's Clinic, University of New South Wales, Sydney, New South Wales 2010, Australia. ar_adji@optusnet.com.au
Objective:
To evaluate the clinical use of radial and carotid artery applanation tonometry as an independent supplement to cuff sphygmomanometry.
Methods:
In 44 patients, radial and carotid tonometric pressure recordings were taken at short intervals apart by two persons who had prolonged experience with both. Comparisons were made between directly recorded radial and carotid waveforms and between aortic waves synthesized from both, using SphygmoCor. Focus was on waveform features: time intervals between wavefoot and incisura, denoting ejection duration, between wavefoot and first systolic peak or shoulder T1, and augmentation index - the rise in pressure from this point to systolic peak divided by pulse pressure.
Results:
No patient had discomfort with radial tonometry, whereas many found carotid tonometry uncomfortable. Beat-to-beat variability was lower for the radial than carotid site. The device's operator "quality index" was achieved for 78% of radial waveforms but just 20% of carotid waveforms (P<0.05). Interobserver variability was lower for all indices derived from radial, cf. carotid, waveforms. For the two observers combined, there was no difference between aortic indices determined from carotid and radial sites except for T1 (radial-derived 117+ or -17 ms, cf. carotid-derived 103+ or -17 ms, P<0.05), but this did not influence the value of augmentation index (radial-derived 26+ or -13%, cf. carotid-derived 28+ or -14%, P=NS).
Conclusion:
The present study conforms with most published results, and indicates superiority of radial to carotid tonometry in clinical practice.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse Assessment Sites
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: