Related Experiment Video
Updated: Aug 7, 2026

Plasmonic Photothermal Cancer Therapy: Nanoparticle-embedded Tumor-tissue-mimicking Phantoms for Visualizing Photothermal Temperature Distribution
Published on: May 9, 2025
Monte Carlo simulation of scattered and thermal photoneutron fluences inside a radiotherapy room
A Facure1, A X Da Silva, R C Falcão
1Comissão Nacional de Energia Nuclear, R. Gal. Severiano 90, sala 405, Rio de Janeiro, RJ 22294-900, Brazil. facure@cnen.gov.br
Abstract:
In this work, the behaviour of scattered and thermal photoneutron fluences in a radiotherapy treatment room was investigated by means of Monte Carlo simulation. The MCNP code was used to study the dependence of these neutron fluences on room design, room area and the effect of neutron moderator materials placed on the room walls. The results of the investigation showed a poor agreement between the simulations and empirical approximations, suggesting that the formulae found in the literature can underestimate the neutron flux inside a radiotherapy room.