High-pressure liquid chromatographic method for the determination of patulin in apple butter
Abstract:
Patulin is extracted from apple butter samples with ethyl acetate and the extract is cleaned up on a silica gel column, using benzene-ethyl acetate (75+25) as the eluant. High-pressure liquid chromatography, using a 25 cm ZorbaxSil column, isooctane-ethyl ether-acetic acid (750+250+0.5) as the mobile solvent, and a 254 nm ultraviolet detector, is used for the determinative step. Under these conditions, patulin is eluted before 5-hydroxymethylfurfural, a component of apple butter which interferes with other liquid chromatographic and thin layer chromatographic methods. Recoveries of patulin added at levels of 34.6, 138.4, and 276.8 mug/kg ranged from 89.0 to 112.1%.