Related Experiment Video
Updated: Aug 7, 2026

Microinjectrode System for Combined Drug Infusion and Electrophysiology
Published on: November 13, 2019
A new infusion pathway monitoring system utilizing electrostatic induced potential
Hiromichi Maki1, Yoshiharu Yonezawa, Hidekuni Ogawa
1Department of Clinical Engineering, International Trinity College, Hiroshima 730-0014, Japan.
Abstract:
We have developed a new infusion pathway monitoring system employing linear integrated circuits and a low-power 8-bit single chip microcomputer. The system is available for hospital and home use and it constantly monitors the intactness of the pathway. The sensor is an electro-conductive polymer electrode wrapped around the infusion polyvinyl chloride infusion tube. This records an AC (alternating current) voltage induced on the patient's body by electrostatic coupling from the normal 100 volt, 60 Hz AC power line wiring field in the patient's room. If the injection needle or infusion tube becomes detached, then the system detects changes in the induced AC voltage and alerts the nursing station, via the nurse call system or PHS (personal handy phone System).
Related Concept Videos
Capillary Electrophoresis: Instrumentation
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...