Related Experiment Video
Updated: Aug 7, 2026

In Vitro Method to Control Concentrations of Halogenated Gases in Cultured Alveolar Epithelial Cells
Published on: October 23, 2018
Badge-type diffusive sampler using 3-methyl-2-benzothiazolinone hydrazone for measuring formaldehyde in indoor air
1R&D Center, Envors Co. Ltd., Birae-dong 151-13, Daeduk-gu, Daejeon 306-809, Korea. envsono@hanmail.net
This study validates a badge-type diffusive sampler for formaldehyde detection using MBTH. The sampler demonstrated good precision and accuracy for indoor air quality monitoring, with effective cleaning procedures.
Area of Science:
- Environmental Chemistry
- Analytical Chemistry
- Air Quality Monitoring
Background:
- Formaldehyde is a common indoor air pollutant with health implications.
- Accurate and reliable measurement methods are crucial for assessing indoor formaldehyde levels.
- Badge-type diffusive samplers offer a passive and convenient approach to air sampling.
Purpose of the Study:
- To evaluate the performance of a badge-type diffusive sampler for formaldehyde measurement.
- To assess the impact of cleaning procedures on blank samples.
- To determine the sampling rate, capacity, and detection limits of the sampler.
Main Methods:
- Utilized a badge-type diffusive sampler impregnated with 3-methyl-2-benzothiazolinone hydrazone (MBTH).
- Investigated the effectiveness of cleaning procedures on reducing formaldehyde in blank samples.
- Determined sampling rates, capacity, and limits of detection/quantification under various conditions.
Main Results:
- Cleaning procedures reduced formaldehyde in blanks by ~31% with no significant difference between techniques.
- The sampler exhibited a maximum sampling rate of 22.4 mL/min at 0.05% MBTH.
- Good precision and accuracy were observed for indoor formaldehyde measurements, with LOD/LOQ at 9.7/13.8 ppb for 24h exposure.
- Higher MBTH concentration (0.5%) increased sampler capacity to 3 ppm·h.
Conclusions:
- The badge-type diffusive sampler is a precise and accurate tool for indoor formaldehyde monitoring.
- Optimized cleaning protocols and MBTH concentration can enhance sampler performance and capacity.
- The validated method provides reliable data for indoor air quality assessments.
More Related Videos
10:29Calibrated Passive Sampling - Multi-plot Field Measurements of NH3 Emissions with a Combination of Dynamic Tube Method and Passive Samplers
Published on: March 21, 2016
08:23Analyzing the Photo-oxidation of 2-propanol at Indoor Air Level Concentrations Using Field Asymmetric Ion Mobility Spectrometry
Published on: June 14, 2018