Evaluation of skin dose in a low dose spiral CT protocol
G Matarese1, M Portelli, M Mazza
1Orthodontics and Pediatric Dentistry Dept., Dental School, University of Messina, Italy.
Aim:
The authors evaluated radiation and skin doses absorbed by the eye lens, as well as the parotid and thyroid glands, during a low dose maxillary spiral computed tomography (CT).
Methods:
Three spiral CTs were carried out, one after the other, changing from time to time MAS, pitch factor, Kv and consequently, CTDI/Vol, until the minimum values detectable by the equipment were reached (MAS 28, Kv 80, CTDI/Vol 2.5, Pitch 1). The quality of the images was evaluated on the grounds of being able to visualise the anatomic structures considered. The amount of radiation effectively absorbed by the soft tissue (skin doses) can be measured with the help of dosimeters placed at the level of eye lens, parotid and thyroid glands. The spiral CTs were performed with a Siemens 16 CT sensation machine, using a tissue equivalent Head-Neck RT Humanoid phantom.
Results:
From the results of this study, it appears possible to obtain high quality images, useful for diagnosing numerous skeletal pathologies in orthodontics, reducing of about 90% the radiation dose per unit of volume and, at the same time, reducing the soft tissue (skin) dose of about 85% when compared to pre-established settings.
Conclusion:
The authors conclude their research emphasizing the possibility of applying low dose protocols, not only in orthodontics but in any dental field, whenever a spiral CT is needed for a more detailed diagnosis.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...