Assessment of the short-lived non-pure positron-emitting nuclide (120)I for PET imaging
H Herzog1, S M Qaim, L Tellmann
1Institute of Medicine, Forschungszentrum Juelich, 52425 Juelich, Germany. h.herzog@fz-juelich.de
Purpose:
The non-pure positron-emitting iodine isotope (120)I (T(1/2)=81 min) is a short-lived alternative to (124)I. (120)I has a positron abundance more than twice that of (124)I and a maximum positron energy of 4 MeV. This study was undertaken to evaluate and characterise the qualitative and quantitative PET imaging of (120)I.
Methods:
(120)I was produced via the (120)Te(p,n) reaction on highly enriched (120)Te. The measurements were done with the Siemens scanner HR+ and the 2D PET scanner GE PC4096+. A cylinder containing three cold inserts and a phantom resembling a human brain slice were used to evaluate half-life, positron abundance and background correction. To analyse the image resolution, a -mm tube placed in water was filled with (120)I and (18)F. Comparisons with (18)F, (124)I and (123)I (measured with SPECT) were made using the Hoffman 3D brain phantom.
Results:
The half-life of 81.1 min was reproduced by the PET measurements. The PET-based positron abundance ranged from 47.9% to 55.0%. The reconstructed image resolution found with the HR+ was 5.4 mm FWHM (12.3 mm FWTM), in contrast to 4.6 mm (8.6 mm) when using (18)F. Erroneous positive and negative numbers of radioactivity found in the cold inserts became nearly zero when the background of gamma-coincidences was corrected for. Images of the Hoffman phantom were inferior to those obtained when (18)F or (124)I was applied but superior to the (123)I-SPECT images.
Conclusion:
Our data show that (120)I of high radionuclidic purity can be regarded as a suitable nuclide for the PET imaging of radioiodine-labelled pharmaceuticals.
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...