Related Experiment Video
Updated: Aug 7, 2026

Measurement of Extracellular Ion Fluxes Using the Ion-selective Self-referencing Microelectrode Technique
Published on: May 3, 2015
Ion-selective electrode for measuring low Ca2+ concentrations in the presence of high K+, Na+ and Mg2+ background
Iwona Bedlechowicz-Sliwakowska1, Peter Lingenfelter, Tomasz Sokalski
1Department of Chemistry, Warsaw University, Pasteura 1, Warsaw, Poland. mmajzur@alfa.chem.uw.edu.pl
Abstract:
In this work, ion-selective electrodes for calcium ion were investigated. Two ionophores were used in the membranes: ETH 1001 and ETH 129. An internal filling solution buffered for primary ion was used that allowed the lower detection limit to be decreased down to 10(-8.8) M. Theoretical and experimental electrode characteristics pertaining to both primary and interfering ions are discussed. Better behavior was obtained with the electrode prepared with ETH 129 in the membrane. This electrode would be the most likely candidate for obtaining a low Ca(2+) detection limit in measurements performed with high K(+), Na(+), Mg(2+) background, which is found inside the cells of living organisms, for example. The potentiometric response of the electrode in solutions containing main and interfering ions is in good agreement with simulated curves obtained using the Nernst-Planck-Poisson (NPP) model.
Related Concept Videos
Electrodeposition
Electrodeposition can...
Potentiometry: Membrane Electrodes
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential ensures...
Qualitative Analysis
For instance, group IV...
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in the...
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...