Precision and accuracy of two blood glucose meters: FreeStyle Flash versus One Touch Ultra
Shannon M Rivers1, Michael P Kane, Gary Bakst
1Department of Pharmacy Practice, Albany College of Pharmacy, Albany, NY 12208, USA.
Purpose:
The precision and accuracy of two blood glucose meters were evaluated using finger and forearm blood samples.
Methods:
Duplicate blood glucose measurements on the same forearm and finger as venipuncture were performed with the FreeStyle Flash and the One Touch Ultra. Accuracy was assessed by error-grid analysis and the number of values within 10% of the laboratory reference value. Precision was determined by calculating the absolute mean percent differences in glucose values between the first and second fingers and forearm test results. Forearm testing success was defined as an accurate glucose reading obtained with one lance.
Results:
A total of 100 patients completed the study; 93% had diabetes and 53% were female. Patients' mean +/- S.D. age was 63 +/- 12 years, and glucose measurements ranged from 69 to 354 mg/dL. All finger-stick samples fell within error-grid zones A and B; 72% and 57% of FreeStyle Flash and One Touch Ultra values fell within 10% of the laboratory reference values, respectively (p = 0.027). Forearm samples were successfully obtained in 99 and 74 patients using the FreeStyle Flash and One Touch Ultra (p < 0.001), with 64 and 36 samples, respectively, falling within 10% of the laboratory reference values (p = 0.035). There was no difference in meter precision.
Conclusion:
The FreeStyle Flash and the One Touch Ultra are precise glucose meters; however, the FreeStyle Flash was associated with greater accuracy. Success rates of forearm glucose sampling were significantly greater when the FreeStyle Flash meter was used.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Uncertainty in Measurement: Reading Instruments
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Type II Diabetes Mellitus III: Clinical Manifestations and Diagnosis