Proximal Cadaveric Femur Preparation for Fracture Strength Testing and Quantitative CT-based Finite Element Analysis
Published on: March 11, 2017
[Testing of anisotropic modulus and fluid characteristic of Chinese femur and tibia]
Bo Wang1, Yubo Fan, Mengshi Chen
1School of Architecture and Environment, Sichuan University, Chengdu 610065, China.
Abstract:
The tension moduls and compression modulus of Chinese tibia and femur of male in three different directions, namely axial, circumferential and radial directions, were tested in our experiments. The measured compression moduli are 5.8 +/- 0.34, 3.4 +/- 0.17, 2.1 +/- 0.18 GPa respectively for femur, and 6.1 +/- 0.36, 3.5 +/- 0.17, 2.1 +/- 0.18 GPa respectively for tibia; the measured tension moduli are 15.6 +/- 1.21, 9.8 +/- 0.73, 5.4 +/- 0.52 GPa respectively for femur, and 16.7 +/- 1.09, 10.3 +/- 0.70, 5.6 +/- 0.49 GPa respectively for tibia. The tension and compression ultimate strengths of femur and tibia were tested, too. The measured compression ultimate strengths are 150 +/- 9.3, 111 +/- 6.5, 110 +/- 7.1 MPa respectively for femur, and 191 +/- 12.3, 140 +/- 6.3, 123 +/- 8.3 MPa respectively for tibia; the measured tension ultimate strengths are 115 +/- 9.6, 82 +/- 6.8, 78 +/- 7.3 MPa respectively for femur, and 143 +/- 8.7, 110 +/- 8.3, 97 +/- 6.4 MPa respectively for tibia. Furthermore, the stress-relaxation experiments were conducted. The relation between loading and time during the experiments were obtained by the method of curve fitting. They meet the logarithm function: F = A * B(log t); A is the initial load, and B is the fitting material parameter. For femur, B is equal to 0.985; for tibia, B is equal to 0.987.