Objective measurement of intraocular lens movement and dioptric change with a focus shift accommodating intraocular
Joanne Hancox1, David Spalton, Catherine Heatley
1Ophthalmology Department, St. Thomas' Hospital, London, United Kingdom. johancox@doctors.org.uk
Purpose:
To objectively measure the shift and refractive change of the 1CU accommodating intraocular lens (IOL) (HumanOptics) and compare them to that of a monofocal AcrySof MA30 IOL (Alcon Laboratories) in the fellow eye.
Setting:
Ophthalmology Department, St. Thomas' Hospital, London, United Kingdom.
Methods:
Thirty patients who had bilateral cataract surgery 18 to 24 months previously with a 1CU IOL prospectively randomly allocated to 1 eye and an AcrySof MA30 monofocal IOL to the other eye were examined. Distance correction, near vision, reading fluency, near point, and defocus to minus spheres were measured. Intraocular lens shift to an accommodative stimulus following instillation of pilocarpine 4% was measured with the ACMaster. Refractive change between distance and near was measured with the Tracey wavefront analyzer.
Results:
Of the original 30 patients recruited, complete measurements could only be obtained for 20. There was no significant difference in near visual function with either IOL. A small anterior movement of the 1CU was seen with accommodation 0.010 mm +/- 0.028 (SD). After pilocarpine 4% instillation, a forward movement of 0.220 +/- 0.169 mm was seen with the 1CU compared to a backward movement of 0.028 +/- 0.095 with the MA30. There was no significant correlation between distance corrected near visual acuity and IOL movement. No change in spherical equivalent between distance and near was seen on wavefront analysis of either IOL.
Conclusions:
Small forward movement of the 1CU IOL was seen with accommodation and increased following pilocarpine, compared to the posterior movement of the MA30 IOL. The amount of the IOL shift was not sufficient to provide useful near vision, but the difference suggests that the engineering concept behind the 1CU IOL is valid.