Pulmonary embolism: radiation dose with multi-detector row CT and digital angiography for diagnosis
Emmanuel Coche1, Stefaan Vynckier, Michelle Octave-Prignot
1Departments of Medical Imaging and Radiation Therapy, Cliniques Universitaires St-Luc, Université Catholique de Louvain, Avenue Hippocrate, 10, 1200 Brussels, Belgium. coche@rdgn.ucl.ac.be
Purpose:
To compare radiation dose delivered at four- and 16-detector row computed tomography (CT) with a dose-modulation program and that delivered at digital angiography for evaluation of pulmonary embolism (PE).
Materials And Methods:
The part of the study involving patients (seven women, four men; mean age, 62 years +/- 16 [standard deviation]; range, 41-85 years) was approved by the institutional review board. Patients gave written informed consent. Exposure was performed with an anthropomorphic phantom with thermoluminescent dosimeters for four-detector row CT without the dose-modulation program and 16-detector row CT without and with the dose-modulation program with standard protocols for pulmonary CT angiography (120 kV, 144 mAs, four and 16 detector rows with 1.00- and 0.75-mm section thickness, respectively). Digital angiograms were acquired with four standard projections at 80 kV. For digital angiography, radiation dose was calculated according to phantom measurements and adapted to acquisition and fluoroscopy times. Distribution of dose was compared for CT and digital angiography.
Results:
During pulmonary CT angiography, mean radiation dose delivered at middle of chest was 21.5, 19.5, and 18.2 mGy for four-detector row CT and for 16-detector row CT without and with dose-modulation program, respectively. At the same level, a mean dose of 91 mGy was delivered with digital angiography. The dose adjusted to clinical conditions was 139.0 mGy for digital angiography and could be reduced after technical adjustment. Ratios of maximum dose to mean dose were 1.15 and 2.96 for CT and digital angiography, respectively. With application of the dose-modulation program at 16-detector row CT, radiation dose was reduced 15%-20% at the upper chest.
Conclusion:
Multi-detector row CT delivers a lower radiation dose, with better spatial distribution of dose, than does pulmonary digital [corrected] angiography. With 16-detector row CT and a dose-modulation program, radiation dose is decreased during PE work-up.
More Related Videos
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Pulmonary Embolism II: Diagnostic Studies and Interprofessional Care
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...