The determination of minoxidil in human serum by high-performance liquid chromatography with amperometric detection
1Control Analytical Research and Development, The Upjohn Company, Kalamazoo, MI 49001, USA.
Abstract:
An ion-pairing reversed-phase HPLC assay employing amperometric detection has been developed for the determination of free minoxidil (MNX) in human serum. The drug is isolated from the serum and concentrated by solid phase extraction with a disposable cartridge column containing ethylsilane (C2) bonded phase packing material. The average absolute recovery from serum was 85%. The HPLC separation is performed on a Sperisorb-C(8) column, using n-octanesulphonic acid as the pairing ion. The method exhibits linear behaviour from 0.3 to 100 ng/ml for spiked serum samples. The average daily relative standard deviations of replicate samples at the 0.75 and 2.0 ng/ml levels were 9.6 and 6.4%, respectively. Utilizing a 1 ml sample the limit of detection (S/N > or = 3) was 0.3 ng/ml.
More Related Videos
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
08:41Integration of Miniaturized Solid Phase Extraction and LC-MS/MS Detection of 3-Nitrotyrosine in Human Urine for Clinical Applications
Published on: July 14, 2017