Related Experiment Video
Updated: Aug 6, 2026

Employing Digital Droplet PCR to Detect BRAF V600E Mutations in Formalin-fixed Paraffin-embedded Reference Standard Cell Lines
Published on: October 8, 2015
Off-line automated sample preparation--experience in a clinical drug investigation unit
S J McIntosh1, I D Watson, P W Hughes
1Drug Investigation Unit, Department of Biochemistry, Royal Infirmary, Glasgow G4 0SF, UK.
Abstract:
The authors describe an evaluation of the Du Pont Prep. off-line automatic centrifugal sample processing system, used in the authors' Drug Investigation Unit over the past two years. In comparison to liquid-liquid extraction, the authors found that assay precision has generally been improved, absolute recovery was quantitative, reliability has been good and the use of the system has resulted in notable labour saving. However, the cost per test based on a seven-year amortization of capital cost shows that the system is more expensive. Assays for urinary opiates, serum tricyclic antidepressants, serum isoxicam, serum valproate, serum metoclopramide and serum pindolol are presented and the merits of the method are compared with those for manual liquid-liquid extraction procedures.