Contrast-detail analysis of three flat panel detectors for digital radiography
Giovanni Borasi1, Ehsan Samei, Marco Bertolini
1Servizio di Fisica Sanitaria, Azienda Ospedaliera Santa Maria Nuova, V.le Risorgimento 80, 42100 Reggio Emilia (RE), Italy. gianni.borasi@asmn.re.it
Abstract:
In this paper we performed a contrast detail analysis of three commercially available flat panel detectors, two based on the indirect detection mechanism (GE Revolution XQ/i, system A, and Trixell/Philips Pixium 4600, system B) and one based on the direct detection mechanism (Hologic DirectRay DR 1000, system C). The experiment was conducted using standard x-ray radiation quality and a widely used contrast-detail phantom. Images were evaluated using a four alternative forced choice paradigm on a diagnostic-quality softcopy monitor. At the low and intermediate exposures, systems A and B gave equivalent performances. At the high dose levels, system A performed better than system B in the entire range of target sizes, even though the pixel size of system A was about 40% larger than that of system B. At all the dose levels, the performances of the system C (direct system) were lower than those of system A and B (indirect systems). Theoretical analyses based on the Perception Statistical Model gave similar predicted SNRT values corresponding to an observer efficiency of about 0.08 for systems A and B and 0.05 for system C.
More Related Videos
07:013D Imaging of Soft-Tissue Samples using an X-ray Specific Staining Method and Nanoscopic Computed Tomography
Published on: October 24, 2019
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Determination of Crystal Structures
Three-Dimensional Microscopy in Microbiology