A robust liquid chromatography tandem mass spectrometry method for total plasma homocysteine determination in
Véronique Ducros1, Hélène Belva-Besnet, Bruno Casetta
1Département de Biologie Intégrée, Centre Hospitalier Universitaire, Grenoble, France. vducros@chu-grenoble.fr
Background:
Total plasma homocysteine has emerged as an independent risk factor for vascular disease. To meet increasing requests by clinicians for this homocysteine determination, a rapid assay for a routine use has been developed.
Methods:
A robust, stable, isotope-dilution liquid chromatography tandem mass spectrometry (LC/MS/MS) method is described, including all the practical details and analytical performance results.
Results:
The method allows homocysteine quantitation over a linear working range up to 100 micromol/L, with the limit of quantification estimated at a low value of 0.09 micromol/L. Total analytical imprecision is less than 4%. Accuracy was assessed by measuring the homocysteine concentration in a serum Standard Reference Material.
Conclusions:
The method was demonstrated to be quick, reliable and cheap after 1 year of use on a time-shared instrument in our hospital unit.
More Related Videos
07:25Quantitative Metabolomics of Saccharomyces Cerevisiae Using Liquid Chromatography Coupled with Tandem Mass Spectrometry
Published on: January 5, 2021
08:56Quantitative Analysis of the Cellular Lipidome of Saccharomyces Cerevisiae Using Liquid Chromatography Coupled with Tandem Mass Spectrometry
Published on: March 8, 2020
Related Concept Videos
Tandem Mass Spectrometry
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process: