Background:
In some children with ventricular rate responsive demand (VVIR) pacemakers (PM), transvenous leads fail for technical reasons or patient's growth.
Aim:
The aim of this study is to describe our experience in adolescents with a nonfunctional ventricular lead in whom the lead was abandoned and an additional VDD lead was implanted. Of the 136 children who received a VVIR PM with an endocardial lead in our center, seven patients aged 7 (0.3-12) years [median (range)] at initial implantation, after 10 (5-15) years showed lead malfunction and underwent atrial synchronous ventricular inhibited pacing (VDD) PM upgrading at 16 (10-20) years.
Results:
The VDD lead was inserted through the ipsilateral subclavian vein in five patients, the contralateral in two (venous occlusion in one and for operator choice in the first patient). The tip was positioned into the right ventricular apex, the atrial dipole along the lateral atrial wall. Fluoroscopy times were not significantly different from those measured in SSI PM implantation and in VVIR dual-chamber demand pacing (VVIR-DDD) upgrading. There were no intraprocedural complications. Follow-up duration is 12 (6-62) months. The VDD PM showed good function, no undersensing or oversensing. Tricuspid damage, new venous occlusion, and "twisting" of the two leads at x-ray were not documented. The first patient showed an infection of the old PM pocket after 1 year, local pain after 3 years, and endocarditis of the leads after 5 years.
Conclusion:
The upgrading of VVIR PM to VDD PM with the abandonment of the nonfunctional lead is feasible, with no intraprocedural complications and good PM function. Lead endocarditis occurred in one patient.
More Related Videos
05:18Implantation of Left Ventricular Assist Device (LVAD) in Juvenile Landrace Swine: A LVAD Implantation Model of Pediatric Heart Failure
Published on: January 16, 2026
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
Cardiomyopathy IV: Restrictive Cardiomyopathy
Cardiomyopathy III: Hypertrophic Cardiomyopathy
Cardiomyopathy II: Dilated Cardiomyopathy
Mitral Stenosis III: Medical Management
Mitral Regurgitation III: Medical Management