Flow-injection determination of thyroxine using immobilized enzyme with tris(2,2'-bipyridyl)ruthenium(III)
Amir Waseem1, Mohammad Yaqoob, Abdul Nabi
1Department of Chemistry, University of Balochistan, Quetta, Pakistan.
Abstract:
A flow-injection method is reported for the determination of thyroxine based on its enhancement effect on the tris(2,2'-bipyridyl)ruthenium(III) chemiluminescence reaction in the presence of NADH using immobilized alcohol dehydrogenase purified from baker yeast. The limit of detection (3 sigma blank) was 5.0 x 10(-8) M with a sample throughput of 80 h(-1). The calibration graph was linear over the range 0.5 - 10 x 10(-7) M (r2= 0.9988) with the relative standard deviation in the range 1.4 - 3.5% (n = 4). The effect of common excipients used in pharmaceutical preparations, some organic compounds and metal ions was studied. The method was applied to pharmaceutical thyroxine tablets, and the obtained results were not significantly different from the amount quoted.