Related Experiment Video
Updated: Aug 6, 2026

A Single-Channel and Non-Invasive Wearable Brain-Computer Interface for Industry and Healthcare
Published on: July 7, 2023
Comparison of ghosting effects for three commercial a-Si EPIDs
L N McDermott1, S M J J G Nijsten, J-J Sonke
1Department of Radiation Oncology, The Netherlands Cancer Institute-Antoni van Leeuwenhoek Hospital, Amsterdam, The Netherlands.
Abstract:
Many studies have reported dosimetric characteristics of amorphous silicon electronic portal imaging devices (EPIDs). Some studies ascribed a non-linear signal to gain ghosting and image lag. Other reports, however, state the effect is negligible. This study compares the signal-to-monitor unit (MU) ratio for three different brands of EPID systems. The signal was measured for a wide range of monitor units (5-1000), dose-rates, and beam energies. All EPIDs exhibited a relative under-response for beams of few MUs; giving 4 to 10% lower signal-to-MU ratios relative to that of 1000 MUs. This under-response is consistent with ghosting effects due to charge trapping.