Related Experiment Video
Updated: Aug 6, 2026

Coherent anti-Stokes Raman Scattering (CARS) Microscopy Visualizes Pharmaceutical Tablets During Dissolution
Published on: July 4, 2014
Calculation of the penetration depth of X-rays in intact pharmaceutical film-coated tablets by microdiffractometry
Hiroyuki Yamada1, Raj Suryanarayanan
1Department of Pharmaceutics, University of Minnesota, 308 Harvard Street, SE Minneapolis, Minnesota 55455, USA.
Purpose:
(i) Develop a method to calculate the penetration depth of X-rays in intact film-coated tablets, and validate it using model bilayer tablets. (ii) Characterize the physical form of drug in intact pharmaceutical film-coated tablets by XRD.
Materials And Methods:
An equation for the calculation of the penetration depth of X-rays, as a function of the incident angle, was derived. Model bilayer tablets were prepared to validate the calculation method. The upper layer of the tablets consisted only of microcrystalline cellulose, while the lower layer was a mixture of cerium oxide (10% w/w), blue dye (5% w/w) and microcrystalline cellulose. The total tablet thickness was 3,500 microm, with the upper layer thickness ranging from 200 to 700 microm. The diffracted intensity of a cerium oxide peak in the lower layer was determined using a microdiffractometer system (CuKalpha radiation) with a two-dimensional area detector. The calculated penetration depth of X-rays was compared with that determined by XRD. After validation of the XRD method, commercial ibuprofen tablets were characterized.
Results:
The penetration depth calculated by the method developed in this study was, in general, in good agreement with that determined experimentally by XRD. In commercial ibuprofen tablets, the coating material exhibited peaks due to TiO2 (25.4 degrees 2 theta) and Fe2O3 (33.3 degrees). However, these did not interfere with the characteristic peak of ibuprofen (22.2 degrees).
Conclusion:
We developed a method for calculation of penetration depth of X-rays in film-coated tablets and validated it using the model bilayer tablets. This method enables the characterization of the active pharmaceutical ingredient in different regions (at different depths) of the film-coated tablet. Since the technique is nondestructive, the same tablet can be repeatedly analyzed during stability studies.
Related Concept Videos
Factors Influencing Drug Absorption: Pharmaceutical Parameters
X-ray Diffraction of Biological Samples
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal crystal...
In Vitro Drug Dissolution: Compendial Testing Models II
Determination of Crystal Structures
Drug Dissolution: Requirements and Profile Comparison
Noncompartmental Analysis: Mean Transit, Absorption and Dissolution Time
One of the key parameters is the mean transit time (MTT), which refers to the total duration required for drug molecules to transit through the body. MTT is determined by calculating the ratio of the area under the moment curve to the area...