Quantitative analysis of deuterium using laser-induced plasma at low pressure of helium
Koo Hendrik Kurniawan1, Tjung Jie Lie, Maria Margaretha Suliyanti
1Research Center of Maju Makmur Mandiri Foundation, 40 Srengseng Raya, Kembangan, Jakarta Barat 11630, Indonesia. kurnia18@cbn.net.id
Abstract:
It was proved that the analysis of deuterium can be conducted using laser-induced plasma spectroscopy. By selecting the appropriate surrounding gas, its pressure, and gating time of the detection system, it was shown that the emission lines of both hydrogen (H(alpha)) and deuterium (D(alpha)), separated by only 0.179 nm, can be fully resolved. A linear calibration curve was also obtained, indicating that this technique has the potential for quantitative analysis of deuterium. The minimum detection limit achieved in this stage of research was estimated to be 50 ppm. We have also shown that this technique can be used as a simple and rapid method for D and H analysis in solid samples.
More Related Videos
Related Concept Videos
¹H NMR of Labile Protons: Deuterium (²H) Substitution
Atomic Emission Spectroscopy: Lab
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...