Hand volume estimates based on a geometric algorithm in comparison to water displacement.

H N Mayrovitz1, N Sims, C J Hill

  • 1College of Medical Sciences, Nova Southeastern University, Ft Lauderdale, Florida 33328, USA. mayrovit@nova.edu

Lymphology
|August 17, 2006
PubMed
Summary

A new elliptical algorithm accurately estimates hand volumes using caliper measurements, offering a practical alternative to water displacement for lymphedema therapy monitoring.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Assessing Unilateral Arm Lymphedema Status Using Interarm Tissue Dielectric Constant Ratios.

Lymphology·2026
Same author

Anatomical and Physiological Considerations for Pig Cardiac Xenotransplantation.

Xenotransplantation·2025
Same author

Systematic review of the application of the Kidney Failure Risk Equation and Oxford classification in estimating prognosis in IgA Nephropathy.

Systematic reviews·2025
Same author

Exploring the experiences of parents of Autistic children when attending the diagnostic imaging department for an X-ray examination.

Radiography (London, England : 1995)·2023
Same author

Exploring the presence of markers of decidualization in the fallopian tubes: a systematic review.

Biology of reproduction·2023
Same author

Systematic review of wastewater surveillance of antimicrobial resistance in human populations.

Environment international·2022

Area of Science:

  • Biomedical Engineering
  • Clinical Measurement
  • Lymphedema Research

Background:

  • Accurate assessment of upper extremity limb volume is crucial for lymphedema therapy efficacy.
  • Traditional tape measurements may be inaccurate for hand volume due to irregular hand shape.
  • A reliable, non-invasive method for hand volume estimation is needed.

Purpose of the Study:

  • To develop and validate a novel measurement procedure and algorithm for estimating hand volumes.
  • To provide a practical alternative to water displacement for routine clinical use.

Main Methods:

  • A caliper was used to measure hand width and depth in 33 subjects (66 hands).
  • Hand volumes were calculated using an elliptical frustum model (VE).
  • VE was compared to water displacement (VW), tape-measure circumference (VC), and trapezoidal model (VT) volumes using regression and LOA.

Related Experiment Videos

Main Results:

  • The elliptical algorithm (VE) showed high correlation with water displacement (VW) (r=0.986), with minimal difference (VE = 1.01VW - 3.1 ml).
  • Limits of Agreement (LOA) for VE vs. VW were +/- 33.5 ml and +/- 9.9%.
  • Tape-measure (VC) and trapezoidal (VT) models significantly overestimated hand volumes (p<0.0001).

Conclusions:

  • The developed elliptical algorithm provides a reliable and accurate method for estimating hand volumes.
  • This method is a valuable alternative to water displacement, especially when the latter is impractical or contraindicated.
  • The algorithm facilitates routine clinical assessment of hand volume changes in lymphedema management.