A comparison of the Doppler spectra from human blood and artificial blood used in a flow phantom
P R Hoskins1, T Loupas, W N McDicken
1Medical Physics and Medical Engineering Department, Royal Infirmary, Edinburgh.
Ultrasound in Medicine & Biology
|January 1, 1990
Abstract:
A comparison between the Doppler signals from human blood and artificial blood used in a flow phantom is described. The artificial blood used was a suspension of Sephadex particles in a glycerol solution. The Doppler power was measured as a function of Sephadex concentration and found to peak at a concentration of about 40% by volume. The power from blood was less by a factor of 150-250 than the power from Sephadex of a similar concentration. The first and second order statistics of the Doppler spectra from Sephadex were independent of particle concentration, and were very similar to those of spectra from blood.