Related Experiment Videos

Clinical results of Hancock II versus Hancock Standard at long-term follow-up

Carlo Valfrè1, Giulio Rizzoli, Claudio Zussa

  • 1Cardiac Surgery Unit, Cà Foncello Hospital of Treviso, Treviso, Italy.

Insights

The Hancock II bioprosthesis offers significantly improved long-term durability compared to the Hancock Standard model, with reduced structural valve deterioration. This enhanced longevity was observed across various patient age groups in a multi-institutional study.

Area of Science:

  • Cardiovascular Surgery
  • Biomaterials Science
  • Clinical Trials

Background:

  • Bioprosthetic heart valves are crucial for treating valvular heart disease.
  • Long-term structural valve deterioration (SVD) remains a primary limitation of bioprostheses.
  • Comparing different generations of bioprosthetic valves is essential for optimizing patient outcomes.

Purpose of the Study:

  • To compare the long-term structural valve deterioration (SVD) between isolated Hancock Standard and Hancock II bioprostheses.
  • To evaluate survival rates and durability of these two bioprosthetic valve models.

Main Methods:

  • A multi-institutional study involving 714 Hancock Standard and 1293 Hancock II bioprostheses implanted between 1983 and 2002.
  • Long-term follow-up (median 12 years, 14,749 patient-years) with 96% completeness.
  • Survival analyzed using Cox analysis, durability using Weibull analysis, and risk factors balanced with a logistic propensity score model.

Main Results:

  • Hancock II had lower perioperative mortality (6% vs. 12%).
  • Age-adjusted 15-year survival was significantly higher for Hancock II (46% vs. 25%).
  • At 15 years, structural valve deterioration (SVD) hazard was 6% for Hancock II vs. 17.5% for Hancock Standard in patients ≥65 years, and 18% vs. 37% in younger patients.

Conclusions:

  • The Hancock II bioprosthesis demonstrates significantly increased durability compared to the Hancock Standard model.
  • After balancing risk factors, Hancock II bioprostheses showed similar survival but substantially improved long-term structural integrity.
  • These findings support the Hancock II as a superior option for long-term valve replacement.