Related Experiment Video
Updated: Jul 20, 2026

Scanning Light Scattering Profiler (SLPS) Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
Characterizing reflections from intraocular lens implants
R J Landry1, I K Ilev, T J Pfefer
1Division of Physics, Center for Devices and Radiological Health, FDA, Rockville, MD 20850, USA. robert.landry@fda.hhs.gov
Objective:
To develop a test method for characterizing glare from intraocular lenses (IOLs) and to confirm a clinical finding that the haptic insertion in the optic of a three-piece IOL produces extraneous line images.
Method:
The method consists of directing a collimated Gaussian laser beam to various parts of the IOL to be tested in a water-filled model eye. Reflected images produced in the retinal plane are photographed with a digital camera.
Results:
A test method was developed to characterize the source of glare images from IOLs. The test method developed was used to confirm a clinical finding that the haptic insertion in the optic of a three-piece IOL produces extraneous line images.
Conclusions:
The method developed can be used to characterize and pin point the source of extraneous glare images from intraocular lens implants. The haptic insertion in the optic of a three-piece IOL has been identified as a source of line images.