Buttressing angle of the double-plating fixation of a distal radius fracture: a finite element study
Chun-Li Lin1, Yu-Hao Lin, Alvin Chao-Yu Chen
1Department of Mechanical Engineering, Chang Gung University, 259 Wen-Hua 1st Road, Kwei-Shan, Tao-Yuan, 333, Taiwan.
Abstract:
Treatment of a distal radius fracture should consider principles including stable fixation and early motion. The aim of this study was to investigate the biomechanical interactions of plate-fixation angles in the internal double-plating method coupled with various load conditions using non-linear finite element analysis (FEA). A 3D finite element distal radius fracture model with three separation angles (50, 70, and 90 degrees ) between the buttressed L- and straight plates was generated based on computed tomography data. After model verification and validation, frictional (contact) elements were used to simulate the interface condition between the fixation plates and the bony surface. The stress/strain distributions and displacements at the radius end were observed under axial, bending, and torsion load conditions. The simulated results indicated that the bending and torsion increased the stress values more than the axial load. The radius and straight plate stress values decreased significantly with increasing fixation angles for all load conditions. However, the L-plate stress values increased slightly under the bending buckling effect. The displacements at the radius end and strains at the fracture healing interface decreased with increasing fixation angles for axial and torsion conditions but displayed a slight difference for the bending condition. The findings using FEA provide quantitative evidence to identify that much larger plate fixation angles could provide better mechanical strength to establish favorable stress-transmission and prevent distal fragment dislocation.
More Related Videos
Related Concept Videos
Applications of Stress
The...
Temperature Dependent Deformation
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Deformation of Member under Multiple Loadings
In the case of a member with a variable cross-section, the strain is not constant but depends on the position. The deformation of an...
Statically Indeterminate Problem Solving
Saint-Venant's Principle