Blood pressure cuff as a potential vector of pathogenic microorganisms: a prospective study in a teaching hospital
C de Gialluly1, V Morange, E de Gialluly
1Laboratory of Microbiology and Hospital Hygiene, University Hospital, Tours, France. c.degialluly@chu-tours.fr
Objective:
To investigate the potential role of blood pressure (BP) cuffs in the spread of bacterial infections in hospitals.
Design:
A comprehensive, prospective study quantitatively and qualitatively evaluating the bacterial contamination on BP cuffs of 203 sphygmomanometers in use in 18 hospital units from January through March 2003.
Setting:
A university hospital with surgical, medical, and pediatric units.
Results:
A level of contamination reaching 100 or more colony-forming units per 25 cm(2) was observed on 92 (45%) of inner sides and 46 (23%) of outer sides of 203 cuffs. The highest rates of contamination occurred on the inner side of BP cuffs kept in intensive care units (ICUs) (20 [83%] of 24) or on nurses' trolleys (27 [77%] of 35). None of the 18 BP cuffs presumed to be clean (ie, those that had not been used since the last decontamination procedure) had a high level of contamination. Potentially pathogenic microorganisms were isolated from 27 (13%) of the 203 BP cuffs: 20 of these microorganisms were Staphylococcus aureus, including 9 methicillin-resistant strains. The highest rates of contamination with potentially pathogenic microorganisms were observed on cuffs used in ICUs and those kept on nurses' trolleys. For 4 patients with a personal sphygmomanometer, a genetic link was found between the strains isolated from the BP cuffs and the strains isolated from the patients.
Conclusions:
The results of this survey highlight the importance of recognizing BP cuffs as potential vectors of pathogenic bacteria among patients and as a source of reinfection when dedicated to a single patient, emphasizing the urgent need for validated procedures for their use and maintenance.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...