Related Experiment Video
Updated: Jul 20, 2026

A Hydrogen-Deuterium Exchange Mass Spectrometry (HDX-MS) Platform for Investigating Peptide Biosynthetic Enzymes
Published on: May 4, 2020
Use of peptide retention time prediction for protein identification by off-line reversed-phase HPLC-MALDI MS/MS
Oleg V Krokhin1, Stephen Ying, John P Cortens
1Department of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba R3T 2N2, Canada. krokhino@cc.umanitoba.ca
Abstract:
A new algorithm, sequence-specific retention calculator, was developed to predict retention time of tryptic peptides during RP HPLC fractionation on C18, 300-A pore size columns. Correlations of up to approximately 0.98 R2 value were obtained for a test library of approximately 2000 peptides and approximately 0.95-0.97 for a variety of real samples. The algorithm was applied in conjunction with an exclusion protocol based on mass (15 ppm tolerance) and retention time (2-min tolerance for 0.66% acetonitrile/min gradient), MART criteria to significantly reduce the instrument time required for complete MS/MS analysis of a digest separated by RP HPLC. This was confirmed by reanalyzing the set of HPLC-MALDI MS/MS data with no loss in protein identifications, despite the number of virtually executed MS/MS analyses being decreased by 57%.
Related Concept Videos
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
MALDI-TOF Mass Spectrometry