Design and experiment of 256-element ultrasound phased array for noninvasive focused ultrasound surgery
Mingzhu Lu1, Mingxi Wan, Feng Xu
1The Key Laboratory of Biomedical Information Engineering of Ministry of Education, Department of Biomedical Engineering, School of Life Science and Technology, Xi'an Jiaotong University, Xi'an 710049, PR China. mzlu@mail.xjtu.edu.cn <mzlu@mail.xjtu.edu.cn>
Abstract:
A 256-element phased array high intensity focused ultrasound system has been designed and constructed in our laboratory. The 256-element spherical-section ultrasound phased array made from piezocomposite material operates at 1.1 MHz with 11-cm radius of curvature, 14-cm outer diameter, and 3.4-cm diameter central hole for mounting diagnostic ultrasound imaging probe. First, the explicit sound field calculation approach for the spherical-section phased array and the genetic optimal algorithm are briefly introduced as the optimal focus pattern control methods. Then, the design guidelines of 256-element spherical-section focused ultrasound phased array and 256-channel driver system are given. The results of single on and off axial focus, multiple on and off axial foci, half sub-array focus pattern provide further evidence for the 3D focus steering and sub-array mode for avoiding obstacles in focused ultrasound surgery. The multi-foci pattern can enlarge the treatment volume to 22 times larger than that of a single focus in one sonication. Finally, in vitro and transparent tissue-mimicking phantom experiment results confirm the ability of 256-element spherical-section phased array system to induce thermal lesions for noninvasive ultrasound surgery.
More Related Videos
05:56Evaluation of the Feasibility, Safety, and Accuracy of an Intraoperative High-intensity Focused Ultrasound Device for Treating Liver Metastases
Published on: January 9, 2019
07:38Real-time Monitoring of High Intensity Focused Ultrasound (HIFU) Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound (HMIFU)
Published on: November 3, 2015