Related Experiment Videos
Mechanical bond strength of the cement-tibial component interface in total knee arthroplasty
Gavin T Pittman1, Christopher L Peters, Jerod L Hines
1Department of Orthopaedics, University of Utah School of Medicine, Salt Lake City, Utah, USA.
Abstract:
The purpose of this study was to mechanically test the cement-tibial component interface using titanium and cobalt-chrome sample prostheses with several commercially available surface textures. The results of this study indicate that the type of metal substrate and surface preparation of contemporary tibial baseplates may influence the strength of the metal-cement interface and as such influence tibial component survival. The results indicate that, in general, metal-cement interface strength increases with increasing surface roughness and common surface treatments such as AlO2 grit-blasting (Ra = 6.76 microm) produce interface strengths similar to plasma-spray, porous-coated specimens. Macrosurfaced tibial components, although comparable in tension, may be vulnerable to metal-cement interface failure with rotational loading.
Related Concept Videos
Bonding and Strength of Aggregate
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris group...
Bones of the Lower Limb: Tibia and Fibula