Reduced burden of bacterial airway colonization with a novel silver-coated endotracheal tube in a randomized
Jordi Rello1, Marin Kollef, Emili Diaz
1Critical Care Department, Joan XXIII University Hospital, Tarragona, Spain. jrello.hj23.ics@gencat.net
Objective:
To assess the feasibility and safety of the respiratory infection control (RIC) device, a silver-coated endotracheal tube, and its effect on bacterial burden in the airways.
Design:
Prospective, randomized, single-blind, multiple-center study.
Setting:
Three hospitals in Spain and one in the United States.
Patients:
Patients were eligible adults who required mechanical ventilation for > or =24 hrs and did not have respiratory infections. One hundred forty-nine patients were intubated and analyzed for safety; 121 were intubated > or =24 hrs and analyzed for tube colonization; 67 had negative baseline quantitative endotracheal aspirates and were analyzed for quantitative endotracheal aspirates colonization.
Interventions:
Intubation with the RIC device or a control endotracheal tube.
Measurements And Main Results:
The RIC device was associated with delayed colonization on the tube compared with the control device at the threshold of ++, +++, or > or =10(4) colony-forming units/mL (p = .02, log-rank test; p = .10, Wilcoxon's test) and in quantitative endotracheal aspirates at > or =10(6) colony-forming units/mL (p = .08, log-rank test; p = .05, Wilcoxon's test). The RIC device was associated with reduced colonization rate by days on the tube (p = .04, Wilcoxon's test) and in quantitative endotracheal aspirates (p = .05, Wilcoxon's test) at the same thresholds. The RIC device was associated with lower maximal bacterial burden in tracheal aspirates for 7 days (mean log-transformed burden, 4.2 +/- 2.3 vs. 5.5 +/- 1.7 log colony-forming units/mL; p = .02, Wilcoxon's test). Other between-group differences favored the RIC device but were not significant. Five adverse events were considered device related, including three events in the RIC group and two events in the control group.
Conclusions:
In this prospectively planned, preliminary analysis, the RIC device was feasible and well tolerated. Larger studies are needed to determine whether delayed colonization, reduced colonization rate, and decreased bacterial burden will decrease the incidence of ventilator-associated pneumonia.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Microbiota of the Respiratory Tract