External quality assessment of HbA(1c) for point of care testing

Mathias Bjuhr1, Christian Berne, Anders Larsson

  • 1Section of Clinical Chemistry, Department of Medical Sciences, Uppsala University Hospital, Uppsala, Sweden.

Insights

Point-of-care Hemoglobin A1c (HbA1c) testing using Bayer DCA 2000 devices shows acceptable imprecision and agreement with central laboratory methods. The external quality assurance program for these devices is cost-effective, costing approximately €150 per instrument annually.

Area of Science:

  • Clinical Chemistry
  • Point-of-Care Testing
  • Quality Assurance

Background:

  • Point-of-care (POC) testing for Hemoglobin A1c (HbA1c) is crucial for diabetes management.
  • The Swedish analytical goal for HbA1c testing is a coefficient of variation (CV) below 3%.

Purpose of the Study:

  • To evaluate the long-term imprecision of HbA1c testing using Bayer DCA 2000 instruments in Uppsala county.
  • To assess the cost-effectiveness of an external quality assurance (QA) program for these POC devices.

Main Methods:

  • A 3-year (2002-2004) split-sample analysis was conducted.
  • Patient samples were analyzed on Bayer DCA 2000 POC devices and re-analyzed using a central laboratory HPLC system.
  • Each instrument was tested approximately 8 times annually.

Main Results:

  • The median CV between the HPLC method and the POC instruments was slightly above the 3% target.
  • Despite this, the DCA 2000 systems demonstrated acceptable imprecision and agreement with the central laboratory.

Conclusions:

  • Bayer DCA 2000 systems provide acceptable HbA1c test results with good agreement across different testing locations within the county.
  • The cost of the external QA program was calculated at approximately SEK 1340 (Euro 150) per instrument, indicating a reasonable investment.
Abstract