Attenuation correction for myocardial perfusion imaging. A comparison between SPECT and PET imaging by polar map
1Medical University of Vienna, Währinger Gürtel 18-20, 1090 Vienna, Austria. senta.graf@meduniwien.ac.at
Aim:
We investigated the impact of photon attenuation in myocardial perfusion imaging with SPECT and PET in patients with coronary artery disease. In fact, the regional tracer distribution can be quantitatively assessed by polar map analysis if the effects of photon attenuation are accounted for. PET imaging permits accurate measurement of and correction for photon attenuation, whereas results of attenuation correction in SPECT imaging have been inconsistent.
Patients, Methods:
We compared photon attenuation in resting perfusion imaging studies with SPECT ((99m)Tc-sestamibi) and PET ((13)N-ammonia) from 21 patients. Transaxial images were reconstructed with and without attenuation correction and reoriented into short axis images. Polar map analysis was utilized to generate regional tracer uptake in six anatomical segments.
Results:
Average segmental photon attenuation calculated as the ratio of counts in corrected and uncorrected images was 7.2 +/- 1.4 in SPECT and 14.0 +/- 3.1 in PET imaging (p < 0.01). This attenuation factor was significantly related to body mass index for both methods (p < 0.001). While attenuation correction for SPECT imaging did compensate for attenuation effects in the inferior wall (from -15% to +6% vs. PET), relative tracer uptake in the anterior wall in SPECT images was significantly reduced after attenuation correction (from -2% to -18% vs. PET, p < 0.01).
Conclusion:
Differential effects of attenuation correction for myocardial SPECT perfusion imaging need to be considered when algorithms designed to compensate effects of photon attenuation in SPECT imaging are employed in clinical practice.
More Related Videos
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET