[Random comparative study between intrauterine device Multiload Cu375 and TCu 380a inserted in the postpartum period]
Roger Lara Ricalde1, Gerardo Menocal Tobías, Carlos Ramos Pérez
1Departamento de planificación familiar del Instituto Naciónal de Perinatología. Lomas Virreyes, México. lararoger@yahoo.com.mx
Ginecologia Y Obstetricia De Mexico
|September 15, 2006
Summary
The Multiload Cu375 intrauterine device (IUD) is as safe and effective as the TCu 380A IUD when inserted postpartum. This study found comparable continuation rates and no significant differences in safety outcomes between the two IUDs.
Area of Science:
- Reproductive Health
- Contraception
- Medical Devices
Background:
- Postpartum insertion of intrauterine devices (IUDs) offers a convenient contraceptive option.
- Comparing the safety and effectiveness of different IUD models is crucial for clinical decision-making.
Purpose of the Study:
- To evaluate and compare the safety and effectiveness of the Multiload Cu375 (MLCu 375) and TCu 380A intrauterine devices when inserted during the postpartum period.
Main Methods:
- A randomized comparative study involving 157 patients who received either an MLCu 375 or TCu 380A IUD.
- IUDs were inserted at various postpartum timings: within 10 minutes after vaginal delivery, during cesarean section, or 10-48 hours postpartum.
- Ultrasound assessments and 12-month follow-up visits were conducted to evaluate IUD position, expulsion, removal, and continuation rates.
Main Results:
- One-year expulsion rates were 10.4% for MLCu 375 and 7.7% for TCu 380A, with no significant influence from insertion timing or IUD position.
- Removal rates for bleeding and pain were similar between groups (4.9% and 4.8%).
- Continuation rates at one year were 77.1% for MLCu 375 and 82.6% for TCu 380A, with no statistically significant differences observed.
Conclusions:
- The Multiload Cu375 intrauterine device demonstrates comparable safety and effectiveness to the TCu 380A when inserted postpartum.
- Both IUD models are suitable options for postpartum contraception, with similar expulsion, removal, and continuation rates.