Related Experiment Video
Updated: Jul 20, 2026

Reliable Acquisition of Electroencephalography Data during Simultaneous Electroencephalography and Functional MRI
Published on: March 19, 2021
Nullification of electromagnetic radiation: 50 Hz artifact during electroencephalogram recording
1Singapore Institute of Manufacturing Technology, 71 Nanyang Drive, Singapore 638075, Singapore. murali@SIMTech.a-star.edu.sg
Abstract:
This paper discusses the 50-Hz artifact observed during EEG recording of three patients, using Mindset24 equipment in a biomedical laboratory along with other instruments. The artifact at 50 Hz is created due to electromagnetic (EM) radiation from the environment. Perhaps, the nullification of 50 Hz artifact occurs by the three important means such as (a) grounding of far-field EM current when the patient touches the equipment metal casing, (b) connecting the electrodes in a sequence to obtain differential potentials with suppressed near-field EM radiation potentials of the scalp, and (c) destructive interferences of the EM waves.
More Related Videos
10:23Equipment Setup and Artifact Removal for Simultaneous Electroencephalogram and Functional Magnetic Resonance Imaging for Clinical Review in Epilepsy
Published on: June 23, 2023
06:58Non-restraining EEG Radiotelemetry: Epidural and Deep Intracerebral Stereotaxic EEG Electrode Placement
Published on: June 25, 2016