Detection and validation of the body edge in low count emission tomography images
Leighton R Barnden1, John Dickson, Brian F Hutton
1Department of Nuclear Medicine, The Queen Elizabeth Hospital, 28 Woodville Road, Woodville, SA 5011, Australia. leighton.barnden@nwahs.sa.gov.au
Abstract:
Segmentation of the body edge in tomographic images with low count, noisy edges is needed for both PET and SPECT respiratory motion correction and brain SPECT attenuation correction. To reduce noise we re-projected tomographic images and searched for edges in the projection count profiles or their spatial derivatives. Edge location versus projection angle was fitted with cosine basis functions after rejecting outliers and including information about edges of previous sections. We processed 10 s duration FDG PET of the thorax, HMPAO brain, DAT brain and lung perfusion SPECT. Stable edges for all four types of scan were achieved but with different profiles. Edges were validated against edges of coregistered CT or MRI. The best root mean square (rms) accuracy was 8.2 mm (PET) and 5.2 mm (brain SPECT). Inter-scan variability (standard deviation) in the estimated-to-control edge distance for 17 PET scans was 0.4 mm, for 25 ordered subset expectation maximisation (OSEM) reconstructed brain SPECT 1.0 mm and for 18 filtered back-projection (FBP) reconstructed brain SPECT 1.4 mm.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...