Determination of Cardiac Output in a Porcine Model for Ex Vivo Pulmonary Perfusion
Published on: June 28, 2024
Cardiac output--pulse contour analysis vs. pulmonary artery thermodilution
M Ostergaard1, J Nielsen, J P Rasmussen
1Department of Anaesthesiology, Copenhagen University Hospital, Gentofte, Hellerup, Denmark. morten.ostergaard@dadlnet.dk
Background:
The aims of this study were to determine the agreement between pulmonary artery thermodilution (PA-TD), transpulmonary thermodilution (TP-TD) and the pulse contour method, and to test the ability of the pulse contour method to track changes in cardiac output.
Methods:
Cardiac output was determined twice before cardiac surgery with both PA-TD and TP-TD. The precision (two standard deviations of the difference between repeated measurements) and agreement of the two methods were calculated. Post-operatively, cardiac output was determined with the PA-TD and pulse contour methods, and the bias and limits of agreement were again calculated. Finally, in patients with heart rates below 60 beats/min or a cardiac index of less than 2.5 l/min/m2, atrial pacing was started and the haemodynamic consequences were monitored with the PA-TD and pulse contour methods.
Results:
Twenty-five patients were included. The precisions of PA-TD and TP-TD were 0.41 l/min [95% confidence interval (CI), +/- 0.07] and 0.48 l/min (95% CI, +/- 0.08), respectively. The bias and limits of agreement between PA-TD and TP-TD were - 0.46 l/min (95% CI, +/- 0.11) and +/- 1.10 l/min (95% CI, +/- 0.19), respectively. Post-operatively, the bias and limits of agreement between the PA-TD and pulse contour methods were 0.07 l/min and +/- 2.20 l/min, respectively. The changes in cardiac output with atrial pacing were in the same direction and of the same magnitude in 15 of the 16 patients.
Conclusion:
The precision of cardiac output measurements with PA-TD and TP-TD was very similar. The transpulmonary method, however, overestimated the cardiac output by 0.46 l/min. Post-operatively, cardiac output measurements with the PA-TD and pulse contour methods did not agree, but the pulse contour method reliably tracked pacing-induced changes in cardiac output.
Related Concept Videos
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Pulse
The pulse serves as a clinical indicator...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...