Clinical testing of cellular phone ringing interference with automated external defibrillators
Alexandre Trigano1, Olivier Blandeau, Christian Dale
1Department of Cardiology, Assistance Publique-Hôpitaux de Marseille, Hôpital Nord, 13915 Marseille, France. alexandre.trigano@mail.ap-hm.fr
Cellular phone calls to automated external defibrillators (AEDs) did not disrupt analysis during unshockable rhythms. Only transient speaker noise was observed, with interference prevented at distances over 15 cm.
Area of Science:
- Biomedical Engineering
- Electrophysiology
- Medical Device Safety
Background:
- Automated external defibrillators (AEDs) are critical for treating cardiac arrest.
- The increasing prevalence of mobile devices necessitates evaluating their potential interference with medical equipment.
- Understanding potential electromagnetic interference (EMI) is crucial for patient safety.
Purpose of the Study:
- To investigate the interference potential of cellular phone ringing on AED function.
- To assess if cellular phone calls disrupt AED analysis and defibrillation capabilities.
Main Methods:
- Tested two cellular phone systems: Global System for Mobile Communication (GSM) and Personal Communication Services (PCS).
- Evaluated three AED models (LifePak 20, LifePak 20 P, HeartStart XL) with their respective electrodes.
- Placed cellular phones in three positions relative to the AED during simulated calls and ECG monitoring.
Main Results:
- No analysis dysfunction was observed in AEDs during unshockable rhythms when exposed to cellular phone calls.
- The primary interference noted was transient speaker noise from the AED, occurring before audible ringing.
- This noise interference was preventable when the cellular phone was positioned at least 15 cm away from the AED.
Conclusions:
- Cellular phone calls do not appear to cause significant functional interference with AED analysis during unshockable rhythms.
- The observed speaker noise is a minor, transient issue that can be mitigated by maintaining a safe distance.
- Clinical use of AEDs alongside cellular phones is likely safe, with minimal risk of critical malfunction.