A method for calculation of true coincidence summing correction factors for extended sources
Tim Vidmar1, Matjaz Korun, Branko Vodenik
1Jozef Stefan Institute, Jamova 39, SI-1000 Ljubljana, Slovenia. Tim.Vidmar@ijs.si
Abstract:
In gamma-ray spectrometry, true coincidence summing correction factors for an extended sample can be calculated from full-energy-peak and total efficiencies as if the sample were a point source, if the so-called linear-to-square- (LS) curve, introduced by Blaauw and Gelsema, is known and properly applied. A method is described for obtaining the efficiencies and the corresponding LS-curve for an arbitrary cylindrical sample from calibration measurements in a reference geometry. The approach is aimed at the analysis of samples measured on p-type HPGe detectors in environmental gamma-ray spectrometry and was successfully verified against experimental data.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification (ADCI) and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Extraction: Partition and Distribution Coefficients
For extracting a solute from an aqueous phase into an organic...
Adjusting a Traverse
Interpreting ¹H NMR Signal Splitting: The (n + 1) Rule
The Power Superposition Principle
Distance Corrections
Calibration Curves: Correlation Coefficient