Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
Cyclic fatigue of different nickel-titanium endodontic rotary instruments
Teresa Roberta Tripi1, Antonio Bonaccorso, Guglielmo Guido Condorelli
1Faculty of Dentistry, University of Catania, Catania, Italy.
Objective:
A comparative study of the fatigue resistance of rotary nickel-titanium endodontic instruments was performed with the aim of assessing the influence of both instrument design and surface treatment on flexural fracture.
Study Design:
To evaluate fatigue resistance of different rotary instruments, a total of 120 instruments were tested; these came from different sources: ProFile, RaCe, K3, Hero, and Mtwo. To compare the effect of electro-polishing procedures on fatigue resistance, a group of RaCe instruments (which are normally electro-polished) without surface treatment was used. A scanning electron microscope (SEM) study for each instrument was performed before and after fatigue study to determine the mode of fracture and the aspect of tips and cross-sectional surface areas.
Results:
ProFile instruments gave the best values for fatigue resistance. It was seen that for RaCe instruments the surface treatment reduces the presence of micro-cracks, surface debris, and machining damage.
Conclusions:
This study shows that the instrument design often proves to be an important factor in the fatigue resistance of NiTi rotary instruments. In RaCe instruments the electro-polishing surface treatment increases the fracture-related fatigue resistance.