Exhaled-breath condensate pH can be safely and continuously monitored in mechanically ventilated patients
Brian K Walsh1, Dan J Mackey, Thomas Pajewski
1Department of Pediatrics, University of Virginia, Charlottesville, VA 22908, USA. bkw2j@virginia.edu
Background:
Airway inflammation in acute and chronic respiratory diseases is characterized in part by abnormal pH in airway-lining fluid. The pH of exhaled-breath condensate (EBC) is low (acidic) in various pulmonary inflammatory diseases, including asthma, chronic obstructive pulmonary disease, cystic fibrosis, pneumonia, and acute respiratory distress syndrome. Because the time course of pH changes in the airway is not yet clear, we aimed to develop a method for frequent and intensive EBC pH data collection in mechanically ventilated patients.
Methods:
We examined the collection, gas-standardizing (CO2 removal), and continuous monitoring of pH of EBC from the expiratory port of a Servo-i ventilator with mechanically ventilated patients. We developed a condensing device that attaches to the exhaust port and is chilled by an electric cooling system. We built a 2-chamber gas-standardization and pH-measuring device that attaches to the condensing system and records pH every 6 s. After safety testing, we enrolled mechanically ventilated patients (with diverse reasons for requiring ventilatory support) for up to 96 h of continuous EBC pH condensimetry.
Results:
The pressure, volume, and flow of the ventilator attached to a test lung were unchanged by application of the condensimeter, at various flows (2-120 L/min) and ventilator settings. We monitored 19 pediatric patients for 6-96 h. The pH of the accumulated EBC in the storage container correlated with the geometric mean of all the pH data points from the condensimeter during the recording period (r2 = -0.95, p < 0.001), which internally validated that the condensimetry system provides accurate, well gas-standardized readings for up to 96 h. The EBC pH values were similar to published reports of single samples. The EBC pH became more acidic during clinical deterioration and normalized with recovery.
Conclusion:
Continuous monitoring of EBC pH from the ventilator exhaust port is safely achievable and reliably provides data that may become useful in monitoring critically ill patients.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Acute Respiratory Failure-III
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...