Three-Dimensionally Printed Microfluidic Cross-flow System for Ultrafiltration/Nanofiltration Membrane Performance Testing
Published on: February 13, 2016
Performance of polysulfone membrane dialyzers and dialysate flow pattern
Ayumu Nakashima1, Satoshi Ogata, Shigehiro Doi
1Department of Molecular and Internal Medicine, Graduate School of Biomedical Sciences, Hiroshima University, Hiroshima, and Department of Internal Medicine, Sanyo Hospital, Fukuyama, Japan.
The TS-UL dialyzer demonstrated more homogeneous dialysate flow and higher clearance of substances compared to the APS-S dialyzer. These findings suggest design differences in the dialysate compartment influence dialyzer performance.
Area of Science:
- Nephrology
- Biomedical Engineering
- Medical Imaging
Background:
- Dialysate flow patterns are crucial for evaluating dialyzer function and optimizing design.
- Polysulfone membrane dialyzers (TS-UL and APS-S) were selected for investigation.
- Computed tomography (CT) with barium sulfate contrast was employed to visualize dialysate flow.
Purpose of the Study:
- To investigate and compare dialysate flow patterns in two polysulfone membrane dialyzers.
- To assess the clinical performance differences between the TS-UL and APS-S dialyzers.
Main Methods:
- In vitro: CT scans with barium sulfate contrast visualized dialysate flow in mock blood and dialysate.
- In vitro: Barium sulfate concentration was measured in defined regions of interest.
- In vivo: 12 hemodialysis patients switched from APS-S to TS-UL for 1 month to assess clinical parameter changes.
Main Results:
- CT scans revealed homogeneous dialysate distribution in TS-UL, but not in APS-S.
- Dialysate concentration curves were similar for TS-UL across regions, unlike APS-S.
- Significantly higher urea nitrogen clearance and albumin loss were observed with TS-UL compared to APS-S.
Conclusions:
- The TS-UL dialyzer exhibited superior clearance of substances across a wide molecular weight range compared to the APS-S.
- Differences in dialysate compartment design are likely responsible for the observed performance variations.