Ethyl glucuronide and ethyl sulphate determination in serum by liquid chromatography-electrospray tandem mass
Luca Morini1, Lucia Politi, Alessandra Zucchella
1Department of Legal Medicine and Public Health, University of Pavia, Via Forlanini 12-27100 Pavia, Italy.
Background:
Ethyl glucuronide (EtG) and ethyl sulphate (EtS) are two ethanol metabolites that can be detected in serum up to 8 h after ethanol elimination. Their presence is therefore indicative of recent ethanol consumption in case of delayed sampling after an event (e.g. car crash).
Methods:
A LC-ESI-MS-MS method for the determination of EtG and EtS in serum was developed and validated. Two product ions together with the parent ions were monitored for identification. Pentadeuterated-EtG was used as internal standard.
Results:
Excellent linearity for EtG (from 0.22 to 45 micromol/l) and EtS (from 0.40 to 80 micromol/l) was observed (r(2)>or=0.9998). LOD and LLOQ were 0.04 and 0.20 micromol/l for EtG and 0.08 and 0.40 micromol/l for EtS, respectively. Accuracy (bias) and precision (relative standard deviation), studied at four different quality control levels, were always better than 7%. Matrix effects were found to be negligible. The method was applied to several samples obtained from known alcoholics and social drinkers.
Conclusions:
A sensitive and specific determination of EtG and EtS in serum samples was achieved despite a simple and fast sample preparation. To our knowledge, this is the first fully validated method for the simultaneous determination of the two alcohol metabolites in serum.
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.