Dosimetric verification for intensity-modulated radiotherapy of thoracic cancers using experimental and Monte Carlo
Si Young Jang1, H Helen Liu, Xiaochun Wang
1Department of Radiation Physics, University of Texas M. D. Anderson Cancer Center, Houston, TX 77030, USA.
Purpose:
To investigate the dosimetric accuracy of commercial treatment planning systems used in intensity-modulated radiotherapy (IMRT) for thoracic cancer.
Methods And Materials:
Clinical IMRT plans for lung and esophageal cancers and mesothelioma were used to investigate the accuracy of dose calculations from two commercial treatment planning systems (Pinnacle and Corvus systems). Dose distributions were measured with ion chambers and thermoluminescent dosimeters for individual IMRT fields and composite treatment plans in water phantoms and anthropomorphic phantoms. A Monte Carlo-based system was established to compute three-dimensional dose distributions to compare with the treatment planning system calculations.
Results:
Dose calculations from the Pinnacle system were acceptable within 5% of the local dose or a 5-mm distance-to-agreement for 80% of the points measured with ion chambers, 74% of the points measured with thermoluminescent dosimeters, and 96% of the points compared with the Monte Carlo calculations. For the Corvus system, 89% of the points agreed with the measured dose and 98% agreed with the Monte Carlo calculations. Underestimation of the dose from the treatment planning system was found in the low-dose regions (<50% of the prescribed dose), possibly caused by inadequate modeling of the multileaf collimators.
Conclusion:
The Pinnacle and Corvus dose calculations were acceptable for thoracic IMRT in high-dose regions. Beam modeling is likely the most critical factor for the accuracy of IMRT dose calculations.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Radiation: Applications
The average...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...