A pulsatile pressure waveform is a sensitive marker for confirming the location of the thoracic epidural space
Pamela H Lennox1, Hamed S Umedaly, Raymer P Grant
1Department of Anesthesia, Vancouver General Hospital, University of British Columbia, Vancouver, British Columbia, Canada. pamela.lennox@vch.ca <pamela.lennox@vch.ca>
Objectives:
The purpose of this study was to assess the validity of using a pulsatile, pressure waveform transduced from the epidural space through an epidural needle or catheter to confirm correct placement for maximal analgesia and to compare 3 different types of catheters' ability to transduce a waveform.
Design:
A single-center, prospective, randomized trial.
Setting:
A tertiary-referral hospital.
Participants:
Eighty-one patients undergoing posterolateral thoracotomy who required a thoracic epidural catheter for postoperative pain management.
Interventions:
Each epidural needle and each epidural catheter was transduced to determine if there was a pulsatile waveform exhibited.
Measurements And Main Results:
Sensitivity of the pulsatile waveform transduced through an epidural needle to identify correct placement of the epidural needle and the sensitivity of each catheter type to identify placement were compared. In 79 of 81 cases (97.5%), the waveform transduced directly through the epidural needle had a pulsatile characteristic as determined by blinded observers. In a total of 53 of 81 epidural catheters (65.4%), the transduced waveform displayed pulsations. Twenty-four of 27 catheters in group S-P/Sims Portex (Smiths Medical MD, Inc, St Paul, MN) (88.9%) transduced a pulsatile tracing from the epidural space, a significantly greater percentage than in the other 2 groups (p = 0.02).
Conclusions:
The technique of transducing the pressure waveform from the epidural needle inserted in the epidural space is a sensitive and reliable alternative to other techniques for confirmation of correct epidural catheter placement. The technique is simple, sensitive, and inexpensive and uses equipment available in any operating room.
Related Concept Videos
Measurement of Blood Pressure
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
Pulse Assessment Sites
Pulse
The pulse serves as a clinical indicator...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...