Related Experiment Video
Updated: Jul 19, 2026

Determination of High-affinity Antibody-antigen Binding Kinetics Using Four Biosensor Platforms
Published on: April 17, 2017
High-resolution characterization of antibody fragment/antigen interactions using Biacore T100
Giuseppe A Papalia1, Mark Baer, Kenneth Luehrsen
1Center for Biomolecular Interaction Analysis, University of Utah, Salt Lake City, UT 84132, USA.
Abstract:
A Biacore T100 optical biosensor was used to characterize the binding kinetics of a panel of antigen binding fragments (Fabs) directed against the PcrV protein from Pseudomonas aeruginosa. PcrV protein forms part of the type III secretion system complex of this opportunistic pathogen. We demonstrate that the biosensor response data for each Fab collected from three different surface densities of the antigen could be fit globally to a simple 1:1 interaction model. Importantly, we found that the Fabs with the slowest dissociation rate provided the best protection in cell cytotoxicity studies. To further characterize the Fab interactions, binding data were automatically acquired at different temperatures and under different buffer conditions. The comprehensive characterization of these Fabs shows how Biacore T100 can be used to complement protein therapeutic discovery programs from basic research to the selection of therapeutic candidates.